Background of Chinese Periodic risk Evaluation Report
NMPA issued guidance “Periodic risk evaluation report (PRER) of medical device” on 02,07,2020. It applies to class II and III of medical device. For class I medical device, only filing of PRER internally is essential.
PRER is also called Chinese periodicic safety updated reports (PSUR) because it is very similar to European PSUR required at MDR. In the fact it is a mixture of a part of European PUSR and risk analysis in term of vigilance (passive) and post market surveillance (active).
NMPA want to strength the monitoring of safety of in China approved medical device and monitor post market surveillance in China and worldwide.
For required PRER in China it contains worldwide marketing history, vigilance summary, literature evaluation and risk analysis related to adverse effects, adverse events and recalls.
PRER for overseas products can be
submitted in Chinese or bilingual:
English and Chinese and will be submitted electronically through Chinese agent
annually.
Cronograma del Informe Periódico de Evaluación de Riesgos
El período PRER comienza a partir de la fecha de aprobación en el certificado NMPA y finaliza el mismo día (fecha de finalización) durante los primeros 5 años anuales. El plazo para presentar el PRER es de 60 días a partir de la fecha de finalización del mismo.
How to submit PRERs depending on product classes and how to calculate first 5 years of PRER period, below we give a clear insight.
Frecuencia de PRER a NMPA
Los fabricantes de dispositivos médicos de clase I deberían
- FILE PRER annually (just for intern record) after initial registration (5 times) till next renewal
The manufacturers of class II and III medical device should
- SUBMIT PRER annually after initial registration (5 times) till next renewal
- ARCHIVAR PRER (solo para registro interno) desde la primera renovación (una vez cada 5 años)
Ejemplo de período PRER
El período de vigencia de la PRER comienza siempre a partir de la fecha de aprobación inicial. Una fecha crucial para determinar si la PRER es aplicable es el 1 de enero de 2019. Esto se debe a que la guía de la PRER es válida a partir de esa fecha.
Case 1: If first renewal of certificate occurs already before 01, 01, 2019, there is only need to FILE every 5 years one PRER as intern record, not to submit to NMPA.
Case 2: If the NMPA certificate is dated on e.g. 05, 06, 2018, you should submit:
- PRER in 2019: from 05, 06, 2018 to 04, 06, 2019
- PRER in 2020: from 05, 06, 2019 to 04, 06, 2020
- PRER en 2021: del 05/06/2020 al 04/06/2021
- PRER in 2022: from 05, 06, 2021 to 04, 06, 2022
Fecha de finalización del PRER: 05, 06
En los primeros 4 años, el fabricante debe presentar el PRER antes del 04, 09 (dentro de los 60 días hábiles siguientes a la fecha de finalización del período del PRER).
En el 5.º PRER, la situación es algo compleja. El fabricante debe presentar la solicitud 60 días hábiles antes de la fecha de vencimiento del certificado. Normalmente, el fabricante renueva el producto ante la NMPA con más de 6 meses de antelación a la fecha de vencimiento del certificado. En este caso, el 5.º PRER finaliza 60 días antes de la fecha de vencimiento.
- PRER in 2023: from 05, 06, 2022 to 04, 03, 2023 (=04, 06, 2023 minus 60 days)
- 2024-2029: solo FILE PRER como registro, etc.
Caso 3: Si el certificado NMPA tiene fecha, por ejemplo, del 05/06/2019, deberá presentar:
- PRER in 2020: from 05, 06, 2019 to 04, 06, 2020
- PRER en 2021: del 05/06/2020 al 04/06/2021
- PRER in 2022: from 05, 06, 2021 to 04, 06, 2022
- PRER in 2023: from 05, 06, 2022 to 04, 06, 2023
- PRER in 2024: from 05, 06, 2023 to 04, 03, 2024
If manufacturer want to renew certificate early in 04, 03, 2024
- PERE in 2024: from 05, 06, 2023 to 04. 12. 2023 (04, 03, 2024 minus 60 days)
- Date of first renewal --- + 5 years, FILE PRER as record etc.
With the time scale at
figure 1 you can match your product correspondingly.
Analysis of Periodic Risk Evaluation Chapter
Marketing history
It refers to registration information in which model and variants and in which other countries medical device is also approved.
Medidas de control de riesgos anteriores
Here is control measures mainly of AE and recall asked.
These measures come from worldwide registered AE and recalls. Besides it, NMPA requires whether
- Retirada del certificado del producto (debería ser un tipo de registro de comercialización mencionado anteriormente).
- Cambios en el proceso de fabricación
- Las instrucciones de uso se revisan.
- Se produce un cambio que debe notificarse.
- Clinical re-evaluation due to safety concerns
En resumen, si ocurrieron los hechos mencionados anteriormente, la NMPA debería tener dudas sobre la seguridad de los productos aprobados en China. Por lo tanto, conviene ser precavido al informar. De lo contrario, deberá proporcionar razones específicas del país por las que no se informaron o integraron en el registro de cambios ante la NMPA. Para cambios en el proceso de fabricación y la revisión de las instrucciones de uso en chino, deberá presentar una solicitud oficial ante la NMPA en China.
Vigilancia (evento adverso y retirada del mercado = FSCA)
Chinese PSUR requires report of individual and group adverse events and recalls worldwide. Actually in many countries the adverse events affecting group is regarding to severity of incidents regarded automatically to high risk event to register to authority.
Análisis de riesgos
Aquí nos centramos en las causas y las medidas de control del riesgo derivado de los eventos adversos registrados y las retiradas del mercado.
The risk analysis in Chinese PSUR sounds for some manufacturers not immediately understandable. Actually periodic risk analysis at risk assessment of incident represents an output of post market activities of approved medical device. Adverse events, recalls or observed incidents of similar product at authority’s vigilance site have to be in risk analysis and adverse effects of medical device in literature should be evaluated. Subsequently, control measurement of these risks should be also analysed.
Template of Periodic Risk Evaluation Report
1. Información básica
- PSUR
- Info of legal manufacturer
- Nombre del producto, variante y modelo
- Product composition
- Intended use
- Número de certificación
2. Historia del marketing a nivel mundial
- Country
- Product name, model and variant
- Status of registration (Valid, obsolete)
- Initial approved date
- Desolate date if applicable
3. Medidas de control de riesgos previas
Yes or no answers in relate to terms below, evidence record if applicable
- Retirada del producto aprobado
- Recall
- Cambio significativo en el proceso de fabricación
- Revisión de las instrucciones de uso
- Cambio de diseño
- Reevaluación clínica
4. Worldwide adverse event and recall reporting (individual and group)
- AE reporting number
- Product name, model and variant
- Lugar de celebración de AE
- Hora de AE
- Descripción de AE
- Control Measures of AE
5. Análisis de riesgos
Summary of safety hazards should be evaluated. If some incidents happen, the cause to this AE (performance or clinical features, design requirement, manufacturing process, storage and transport condition, operating environment) has to be analysed. The control measures are also a part of risk analysis to conclude that these hazards have no safety concerns left.
6. Conclusión
The current PRER compared to last periodic of PRER should be analysed. The acceptance of difference should be evaluated.
Servicio de Informes Periódicos de Evaluación de Riesgos del Servicio Chino
In Professional service we guarantee that you pass the final review at authority. We will accompany the delegation of task internallys and communication with your Chinese distributor or Chinese agent.
| Regular | Profesional | |
|---|---|---|
| incógnita | ||
| Instrucciones de la plantilla | No | incógnita |
| No | incógnita | |
| No | incógnita | |
| Gestión de proyectos (becarios y estudiantes en prácticas) | No | incógnita |
| No | incógnita | |
| Precio en EUR | 99 |
Regular service
Servicio profesional
Comparación con el Informe periódico de actualización de seguridad (PSUR) del Reglamento de Dispositivos Médicos (MDR, por sus siglas en inglés) (Artículo 86).
Compared to EU PSUR there are some limitations of Chinese PSUR in terms of state of the art:
- No expiration of Chinese PSUR (so far no explanation about cancellation of certificate or no disstribution of medical device)
- No data of similar product
- No data from official database (as EU register)
| IIa, IIb y III | ||
| Anualmente después de la aprobación inicial del producto (5 años) | ||
| Contenidos clave del informe |
Some thoughts about Periodic Risk Evaluation Report
NMPA has elite tempo to strength registration complexity of medical device since 2014. NMPA is aware that at submission dossier there is no post market information. After issuing guidance of adverse event and recall in 2019, Periodic risk evaluation report is born in 2020.
En realidad, el PSUR europeo para dispositivos médicos copia el PSUR de medicamentos, centrándose en los efectos adversos. Es una especie de mezcla de los requisitos de PMS para dispositivos médicos y una de las normativas vigentes en el MDR europeo. China adoptó rápidamente el PSUR, añadiendo requisitos de análisis de riesgos con el nuevo nombre de Informe periódico de evaluación de riesgos.
Como se muestra en la plantilla, el análisis de riesgos abarca un alcance demasiado amplio relacionado con eventos adversos registrados, retiros del mercado, etc. Recomendamos encarecidamente no informar sobre acciones correctivas y preventivas (CAPA) sensibles, revisiones de las instrucciones de uso (IFU) fuera de China, cambios de diseño, etc. De lo contrario, la NMPA podría preocuparse de que el dispositivo médico presente problemas de seguridad a nivel mundial. En su lugar, proporcione un historial completo de comercialización a nivel mundial, eventos adversos, retiros del mercado y sus causas, así como medidas de control en términos de análisis de riesgos.
Lamentablemente, la NMPA ya está elaborando nuevas directrices sobre la vigilancia posterior a la comercialización, y PMCF las sigue de cerca.
Tendencia de los informes de seguridad social internacionales
· Consolidated template of international PSUR
· Digital tool for automatization of international PSUR
Taiwán
It is called 醫療器材安全性總結報告(Safety Summary Report, SSR). It is very rational that only high risk class III device as stent, Heart valves, artificial joints which are initially registered in Taiwan is applicable for a regular safety report
International
The Medical Devices (Post-market Surveillance Requirements) [1]
(Amendment) (Great Britain), 2025
Reglamento UE 2017/745, MDCG 2022-21 - Guía sobre el informe periódico de actualización de seguridad[2], 2022
EU EMA, Periodic safety update reports (drug)[3]
[1] https://assets.publishing.service.gov.uk/media/67813b39363ac763c8ede498/Medical_devices_periodic_safety_update_report__PSUR__formatting__1_.pdf
[2] https://health.ec.europa.eu/latest-updates/mdcg-2022-21-guidance-periodic-safety-update-report-psur-according-regulation-eu-2017745-december-2022-12-16_en
[3] https://www.ema.europa.eu/en/human-regulatory-overview/post-authorisation/pharmacovigilance-post-authorisation/periodic-safety-update-reports-psurs
Gerente externo de asuntos regulatorios
If your marketing holder either in China or in other markets is disributor without regulatory knowledge, it makes sense to outsurce RA to us.
Annual (self-examination) quality management system report
Otro informe similar es el Informe anual (autoevaluación) del sistema de gestión de calidad por agente chino. Hay más indicadores clave de rendimiento (KPI) que recopilar para todos los dispositivos médicos aprobados en China.
Vigilancia posterior a la comercialización en China
Do Not ignore other activties after product approval in China.
International quality management system
¿Cuáles son los requisitos internacionales en materia de sistemas de gestión de calidad (SGC)? Varían según la frecuencia de los informes, el contenido específico de los mismos y su relación con el certificado del producto.
¿Quieres familiarizarte con otras actividades posteriores a la comercialización y convertirte en un experto en el registro chino y global?
Reserva nuestro curso de formación online. ¡Aún tenemos descuento!






